Cirq Arm System - Model SBS 1.3.2 (Art.No. 56000-01B) and SBS 1.4 (Art.No. 56000-01C) of the CIRQ Arm System (56000 and 56000A)
Recall
- Recall Number
- Z-2925-2020
- Event Number
- 86272
- Firm
- Brainlab AG Olof-palme-str. 9 Munich Germany
- FEI Number
- 3002619595
- Product Code
- OLO
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- August 20, 2020
- Terminated
- May 13, 2021
Description
Cirq Arm System - Model SBS 1.3.2 (Art.No. 56000-01B) and SBS 1.4 (Art.No. 56000-01C) of the CIRQ Arm System (56000 and 56000A)
Holding force of the Positioning Arm can be reduced by prolonged non-use.
Brainlab will be distributing the Medineering Field Safety Notice / Product Notification CAPA-17 to the consignees via email. Instruction for user on how to effectively avoid the situation of risk to occur (i.e. how to avoid the described potential risk), please also refer to the Corrective action by the user section of the Medineering Field Safety Notice / Product Notification CAPA-17. 1. Please identify all Positioning Arms with the specified serial numbers 2. Especially when the arm is freestanding and unsupported, ensure that the Positioning Arm is not above the patient or the user. 3. Avoid unfavorable positions of the Positioning Arm. Examples of these can be found in the instructions or use in the payload test section. 4. Check the secure hold after each arm alignment.
USA: DC, OH, IL OUS: Bulgaria; France; Germany; United Kingdom
21 arms