FDA Recall Terminated

Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMT ET Tube, 7.0mm REF: NVTKIT-C7-5, RxOnly, Sterile, UDI: B006NvtKITc752/$$7062818C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Recall: Z-2901-2020 · Initiated July 21, 2020

Recall

Recall Number
Z-2901-2020
Event Number
86102
Firm
Neurovision Medical Products Inc
FEI Number
1930870
Product Code
ETN
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
July 21, 2020
Terminated
April 30, 2021
Address
2443 Portola Rd, Ventura, CA, 93003-7726

Description

Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMT ET Tube, 7.0mm REF: NVTKIT-C7-5, RxOnly, Sterile, UDI: B006NvtKITc752/$$7062818C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Reason

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Action

Urgent Medical Device Recall Notice, dated July 17, 2020 was sent via email on 7/21/20 and a hard copy via USPS first class mail on 7/28/20-7/29/20 to consignees advising them that Neurovision Medical Products (hereafter NMP) is voluntarily recalling Cobra 1 & 2-Channel EMG Endotracheal Tube, 6 mm, 7 mm, and 8 mm which is used for continuous EMG monitoring and status assessment of the nerves. NMP has initiated this removal due to the potential degradation of the electrode insulation material sooner than the established shelf life. This could make the electrode more susceptible to damage by dental scratching or acute bending. This may interfere with the EMG performance of the device. Failure of the device could result in a delay in the surgical procedure. Malfunction during clinical use should be evident to the clinical user by intermittent or loss of EMG signal and high electrode impedance values. We are aware of one product failure and one complaint associated with the problem. It advises customers to immediately take the following actions: Review the enclosed list of affected lots and product codes. 2. Quarantine product subject to this notice. 3. Cease distribution and use of any product from the affected lots. 4. Advise all applicable departments (including Anesthesia, OR Leadership and Materials Management). 5. Complete and return the enclosed response form as soon as possible: a. Email: [email protected] b. Fax: (877) 330-1727 6. NMP Customer Service will respond with a return authorization (RMA) number and return shipping label for any stock of affected lots that you may have. 7. If you have distributed this product to another location or entity, notify them of this product recall (include a copy of this notice), and contact NMP Customer Service for a return authorization (RMA). The letter stated the firm is working quickly and closely with their manufacturing team to address the issue. In the meantime, they do have an alt

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.

Quantity

7,250