FDA Recall Open, Classified

RayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care.

Recall: Z-2899-2020 · Initiated July 29, 2020

Recall

Recall Number
Z-2899-2020
Event Number
86249
Firm
RAYSEARCH LABORATORIES AB Sveavagen 9 Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 29, 2020

Description

RayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care.

Reason

When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image selected. When starting the task, an error message might be displayed stating: Review files not available.

Action

Only 1 effected consignee who was emailed a field safety notice on 08/05/2020. This letter instructs the user to: Be aware that registration of a treatment image with a corresponding study with multiple series in the same frame of reference might block the offline image review. Verify the correctness of the data in RayCare PACS before using the data for any decision making. -Please educate staff and all users about this workaround. -Inspect your product and identify all installed units with the above software version number(s). -Confirm you have read and understood this notice by replying to the notification email.

Distribution

US Nationwide distribution including in the state of Tennessee.