Aquilion One: Aquilion TSX-301B, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning.
Recall
- Recall Number
- Z-2896-2011
- Event Number
- 58031
- Firm
- Toshiba American Medical Systems Inc
- FEI Number
- 2020563
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 7, 2011
- Posted
- July 29, 2011
- Terminated
- August 17, 2012
- Address
- 2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068
Description
Aquilion One: Aquilion TSX-301B, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning.
Toshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may be broken due to failure of the optical conversion connector, resulting in a break of communication between units.
Toshiba America Medical Systems, Inc. sent an "Urgent Medical Device Correction" letter dated January 4, 2011 to all affected customers. The letter provides the customers with an explanation of the problem and the firms plan to install preventive measures and provided them with a Customer Reply form to send Toshiba. A representative will contact the customers for an appointment to install the measures to their CT system. For question on this recall call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.
(USA) Nationwide Distribution including the states of WI, MO, TN, IA, AL, AZ, TX, and GA.
5