Aquilion One: Aquilion TSX-301A, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning.

Recall: Z-2895-2011 · Initiated January 7, 2011

Recall

Recall Number
Z-2895-2011
Event Number
58031
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
January 7, 2011
Posted
July 29, 2011
Terminated
August 17, 2012
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92781-2068

Description

Aquilion One: Aquilion TSX-301A, 64 System, X-Ray, Tomography. Toshiba American Medical Systems, Inc. A multislice CT system that supports whole-body scanning.

Reason

Toshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may be broken due to failure of the optical conversion connector, resulting in a break of communication between units.

Action

Toshiba America Medical Systems, Inc. sent an "Urgent Medical Device Correction" letter dated January 4, 2011 to all affected customers. The letter provides the customers with an explanation of the problem and the firms plan to install preventive measures and provided them with a Customer Reply form to send Toshiba. A representative will contact the customers for an appointment to install the measures to their CT system. For question on this recall call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.

Distribution

(USA) Nationwide Distribution including the states of WI, MO, TN, IA, AL, AZ, TX, and GA.

Quantity

6