FDA Recall Terminated

Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.

Recall: Z-2888-2016 · Initiated December 29, 2009

Recall

Recall Number
Z-2888-2016
Event Number
75160
Firm
Spiegelberg Gmbh & Co. KG Tempowerkring 4 Hamburg Germany
FEI Number
3002108226
Product Code
GWM
Status
Terminated
Root Cause
Packaging
Initiated
December 29, 2009
Terminated
October 5, 2016

Description

Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.1.63/FV535P Product Usage: lntracranial pressure measuring probes are used to measure intracranial pressure (intracranial pressure, ICP), so the pressure in the head.

Reason

In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.

Action

Spiegelberg Gmbh & Co. KG sent an Urgent Medical Device Recall letter dated December 29, 2009, to all affected customers. Customers were instructed to quarantine and return stored products. Furthermore, they were asked to inform all their customers which have received the affected products and recall the products accordingly. Furthermore, they were asked to tell us about the actions taken by them.

Distribution

Nationwide Distribution to PA only

Quantity

22 units