FDA Recall Terminated

Match Point System The Match Point System is intended to by uses as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-image scans.

Recall: Z-2886-2020 · Initiated July 17, 2020

Recall

Recall Number
Z-2886-2020
Event Number
86113
Firm
Materialise N.V. Technologielaan 15 Heverlee Belgium
FEI Number
3003998208
Product Code
KWS
Status
Terminated
Root Cause
Software Design Change
Initiated
July 17, 2020
Terminated
December 23, 2020

Description

Match Point System The Match Point System is intended to by uses as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-image scans.

Reason

The procedure side indicated in the top header of the Shoulder Case Planning Report incorrectly indicated the procedure side as "Left" for surgical plans for a "right" procedure side.

Action

The firm, materialise, sent Recall letters dated 17 July 2020 informing customers of the issue. This letter informs the customer that Materialise has corrected the Shoulder Case Planning Report in the SurgiCase online system. If you have any questions, contact Global Quality and Regulatory Manager Medical, at 734-259-7017 or email: [email protected].

Distribution

Nationwide distribution to: US including PR and International to UK and Australia.

Quantity

128