FDA Recall Terminated

GE Healthcare Optima CT540. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.Sold under the following product names: BEIJING CT OPTIMA 540, BEIJING OPTIMA, BEIJING OPTIMA 540, BEIJING OPTIMA 540 ASIR, CT BEIJING OPTIMA 540, CT BEIJING OPTIMA 540 ASIR, CT GOLDSEAL OPTIMA 540, CT OPTIMA 540, CT OPTIMA 540 ASIR, CT OPTIMA CT540 TANG BEIJING, DISCOVERY 670 3/8 MEA PRO, HINO OPTIMA 540, HINO OPTIMA 540 ASIR, HINO OPTIMA 540 ASIR, OPTIMA 540, OPTIMA 540 ASIR, OPTIMA 540 MOBILE BEIJING, OPTIMA CT540, OPTIMA CT540 32 SLICES, OPTIMA CT540 ASIR, OPTIMA CT540 BJG SKD RUSSIA, Optima CT540 Gantry Top Asm with Rating Plate,SGB tube and CJ cover, Optima CT540 Gantry Top Asm with Rating Plate,SGB tube and CJ cover for Hino, OPTIMA CT540 TANG, OPTIMA CT540 TANG BEIJING, OPTIMA CT540 TANG HINO

Recall: Z-2872-2018 · Initiated April 27, 2018

Recall

Recall Number
Z-2872-2018
Event Number
80563
Firm
GE Healthcare Japan Corporation Ge Herusukea.Japan Kabu 4 Chome 7-127 Asahigaoka Hino Japan
FEI Number
3002808529
Product Code
JAK
Status
Terminated
Root Cause
Employee error
Initiated
April 27, 2018
Terminated
August 21, 2020

Description

GE Healthcare Optima CT540. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.Sold under the following product names: BEIJING CT OPTIMA 540, BEIJING OPTIMA, BEIJING OPTIMA 540, BEIJING OPTIMA 540 ASIR, CT BEIJING OPTIMA 540, CT BEIJING OPTIMA 540 ASIR, CT GOLDSEAL OPTIMA 540, CT OPTIMA 540, CT OPTIMA 540 ASIR, CT OPTIMA CT540 TANG BEIJING, DISCOVERY 670 3/8 MEA PRO, HINO OPTIMA 540, HINO OPTIMA 540 ASIR, HINO OPTIMA 540 ASIR, OPTIMA 540, OPTIMA 540 ASIR, OPTIMA 540 MOBILE BEIJING, OPTIMA CT540, OPTIMA CT540 32 SLICES, OPTIMA CT540 ASIR, OPTIMA CT540 BJG SKD RUSSIA, Optima CT540 Gantry Top Asm with Rating Plate,SGB tube and CJ cover, Optima CT540 Gantry Top Asm with Rating Plate,SGB tube and CJ cover for Hino, OPTIMA CT540 TANG, OPTIMA CT540 TANG BEIJING, OPTIMA CT540 TANG HINO

Reason

GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.

Action

The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative. GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification.

Distribution

Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN.

Quantity

4,007 devices total (products 1, 2, 3)