GE Healthcare Optima CT660. Sold under the following product names: BEIJING OPTIMA CT660 PRO GT170, BEIJING OPTIMA CT660 PRO GT200, BEIJING OPTIMA CT660S PRO 32 S, CT GOLDSEAL, CT GOLDSEAL OPTIMA 660, CT GOLDSEAL OPTIMA 660 UT-BUC, CT OPTIMA 660 LNG HN P20, CT OPTIMA 660 MID HN P20, CT OPTIMA 660 PRO, CT OPTIMA 660 PRO LONG BJG P2., CT OPTIMA 660 PRO MID, CT OPTIMA 660 PRO MID HN P20, CT OPTIMA 660S LNG HN P20, CT OPTIMA CT660 GT 1700, CT OPTIMA CT660 GT1700, CT OPTIMA CT660 GT1700V, CT OPTIMA CT660 PRO GT1700, CT OPTIMA CT660 PRO LNG P1.5 H, CT OPTIMA CT660 PRO MID P1.5 H, CT OPTIMA CT660 SE GT 1700, CT OPTIMA CT660 SE GT2000, CT660, CT660 ASIR OPT, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT2000 MKE25, OPTI660_A_G_1.5, OPTI660_A_L_2.0, OPTI660_N_G_1.5, OPTI660_N_G_2.0, OPTI660_N_L_2.0, OPTIMA 660 DIVA, OPTIMA 660 LNG HN P20, OPTIMA 660 LNG HN P20, OPTIMA 660 LONG, OPTIMA 660 LONG BJG P2.0, OPTIMA 660 LONG WSO, OPTIMA 660 MID, OPTIMA 660 MID BJG P1.5, OPTIMA 660 MID HN P20, OPTIMA 660 MID HN P20, OPTIMA 660 MID WSO, OPTIMA 660 PRO LGN HN P20, OPTIMA 660 PRO LONG BJG P1.5_, OPTIMA 660 PRO LONG BJG P2, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG WSO, OPTIMA 660 PRO MID BJG P1.5, OPTIMA 660 PRO MID BJG P2.0, OPTIMA 660 PRO MID H, OPTIMA 660 PRO MID HN P20, OPTIMA 660 PRO MID WSO, OPTIMA 660 WSO, OPTIMA 660S 32 OPTIMA 660S 32 SLICE LONG WSO, OPTIMA 660S 32 SLICE MID WSO, OPTIMA 660S LNG HN P20, OPTIMA 660S MID BJG P2.0, OPTIMA 660S PRO 32 SLICE M WSO, OPTIMA 660S PRO LONG BJG P2.0, OPTIMA 660S PRO MID BJG P1.5, OPTIMA 660S WSO, OPTIMA CT 660 FREEDOM, OPTIMA CT660, OPTIMA CT660 2ND EDITION, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR ADV 2ND, OPTIMA CT660 FREEDOM, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG HN2, OPTIMA CT660 FREEDOM LG HN2.5, OPTIMA CT660 FREEDOM MD, OPTIMA CT660 FREEDOM MD BJG2., OPTIMA CT660 GS, OPTIMA CT660 GT 1700 BJG2, OPTIMA CT660 GT 1700 BJG2.5, OPTIMA CT660 GT 1700 HN P2, OPTIMA CT660 GT 1700 HN P2., OPTIMA CT660 GT 1700 HN P2.5, OPTIMA CT660 GT 2000 HN P2.5, OPTIMA CT660 GT1700, OPTIMA CT660 GT1700 MKE2.5, OPTIMA CT660 GT1700 PHASE 1.5, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000 PHASE 1.5, OPTIMA CT660 LITE, OPTIMA CT660 LITE HN P2.5, OPTIMA CT660 MID P1.5 HINO, OPTIMA CT660 MID P20, OPTIMA CT660 MID P2000, OPTIMA CT660 PRO, OPTIMA CT660 PRO GT 2000, OPTIMA CT660 PRO GT1700, OPTIMA CT660 PRO GT1700 ASIR, OPTIMA CT660 PRO GT1700 P20, OPTIMA CT660 PRO GT1700 PHASE, OPTIMA CT660 PRO GT1700 PHASE 1.5, OPTIMA CT660 PRO GT1700V, OPTIMA CT660 PRO GT1700V ASIR, OPTIMA CT660 PRO GT2000, OPTIMA CT660 PRO GT2000 PHASE, OPTIMA CT660 PRO GT2000 PHASE 1.5, OPTIMA CT660 PRO LNG, OPTIMA CT660 PRO LNG P1.5, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LONG BJG P2.0, OPTIMA CT660 PRO LONG P2.0, OPTIMA CT660 PRO MID P1.5, OPTIMA CT660 PRO MID P1.5 H, OPTIMA CT660 PRO MID P1.5 HINO, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700 BJG, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 HN2, OPTIMA CT660 SE GT 2000, OPTIMA CT660 SE GT 2000 BJG, OPTIMA CT660 SE GT 2000 BJG2., OPTIMA CT660 SE GT 2000 BJG2.5, OPTIMA CT660 SE GT 2000 HN2, OPTIMA CT660 SE GT 2000 HN2.5, OPTIMA CT660 SE GT2000 MKE2.5, OPTIMA CT660 SE WITH ASIR, OPTIMA CT660 W/PRO PKG-BR, OPTIMA CT660 W/PRO PKG-LA, OPTIMA CT660-64 USA, OPTIMA CT660S, OPTIMA CT660S, OPTIMA CT660S 32 SLICE GT1700, OPTIMA CT660S 32 SLICE HINO, OPTIMA CT660S GT, OPTIMA CT660S GT 1700 HN2.5, OPTIMA CT660S GT 2000, OPTIMA CT660S GT 2000 BJG2.5, OPTIMA CT660S GT 2000 HN2.5, OPTIMA CT660S GT1700 BJG P2.5, OPTIMA CT660S GT1700 HN P2.5, OPTIMA CT660S PRO 32 HINO, OPTIMA CT660S PRO 32 SLICE, OPTIMA CT660S PRO 32 SLICE GT2, OPTIMA CT660S PRO P1.5 HINO, OPTIMACT660_A_G, OPTIMACT660_A_L, OPTIMACT660_N_G, OPTIMACT660_N_L, REVOLUTION EVO, REVOLUTION EVO WITH GT2000 TABLE (H
Recall
- Recall Number
- Z-2871-2018
- Event Number
- 80563
- Firm
- GE Healthcare Japan Corporation Ge Herusukea.Japan Kabu 4 Chome 7-127 Asahigaoka Hino Japan
- FEI Number
- 3002808529
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- April 27, 2018
- Terminated
- August 21, 2020
Description
GE Healthcare Optima CT660. Sold under the following product names: BEIJING OPTIMA CT660 PRO GT170, BEIJING OPTIMA CT660 PRO GT200, BEIJING OPTIMA CT660S PRO 32 S, CT GOLDSEAL, CT GOLDSEAL OPTIMA 660, CT GOLDSEAL OPTIMA 660 UT-BUC, CT OPTIMA 660 LNG HN P20, CT OPTIMA 660 MID HN P20, CT OPTIMA 660 PRO, CT OPTIMA 660 PRO LONG BJG P2., CT OPTIMA 660 PRO MID, CT OPTIMA 660 PRO MID HN P20, CT OPTIMA 660S LNG HN P20, CT OPTIMA CT660 GT 1700, CT OPTIMA CT660 GT1700, CT OPTIMA CT660 GT1700V, CT OPTIMA CT660 PRO GT1700, CT OPTIMA CT660 PRO LNG P1.5 H, CT OPTIMA CT660 PRO MID P1.5 H, CT OPTIMA CT660 SE GT 1700, CT OPTIMA CT660 SE GT2000, CT660, CT660 ASIR OPT, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT1700 MKE25, OPT CT660 FREEDOM GT2000 MKE25, OPTI660_A_G_1.5, OPTI660_A_L_2.0, OPTI660_N_G_1.5, OPTI660_N_G_2.0, OPTI660_N_L_2.0, OPTIMA 660 DIVA, OPTIMA 660 LNG HN P20, OPTIMA 660 LNG HN P20, OPTIMA 660 LONG, OPTIMA 660 LONG BJG P2.0, OPTIMA 660 LONG WSO, OPTIMA 660 MID, OPTIMA 660 MID BJG P1.5, OPTIMA 660 MID HN P20, OPTIMA 660 MID HN P20, OPTIMA 660 MID WSO, OPTIMA 660 PRO LGN HN P20, OPTIMA 660 PRO LONG BJG P1.5_, OPTIMA 660 PRO LONG BJG P2, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG BJG P2.0, OPTIMA 660 PRO LONG WSO, OPTIMA 660 PRO MID BJG P1.5, OPTIMA 660 PRO MID BJG P2.0, OPTIMA 660 PRO MID H, OPTIMA 660 PRO MID HN P20, OPTIMA 660 PRO MID WSO, OPTIMA 660 WSO, OPTIMA 660S 32 OPTIMA 660S 32 SLICE LONG WSO, OPTIMA 660S 32 SLICE MID WSO, OPTIMA 660S LNG HN P20, OPTIMA 660S MID BJG P2.0, OPTIMA 660S PRO 32 SLICE M WSO, OPTIMA 660S PRO LONG BJG P2.0, OPTIMA 660S PRO MID BJG P1.5, OPTIMA 660S WSO, OPTIMA CT 660 FREEDOM, OPTIMA CT660, OPTIMA CT660 2ND EDITION, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR ADV 2ND, OPTIMA CT660 FREEDOM, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG BJG2.5, OPTIMA CT660 FREEDOM LG HN2, OPTIMA CT660 FREEDOM LG HN2.5, OPTIMA CT660 FREEDOM MD, OPTIMA CT660 FREEDOM MD BJG2., OPTIMA CT660 GS, OPTIMA CT660 GT 1700 BJG2, OPTIMA CT660 GT 1700 BJG2.5, OPTIMA CT660 GT 1700 HN P2, OPTIMA CT660 GT 1700 HN P2., OPTIMA CT660 GT 1700 HN P2.5, OPTIMA CT660 GT 2000 HN P2.5, OPTIMA CT660 GT1700, OPTIMA CT660 GT1700 MKE2.5, OPTIMA CT660 GT1700 PHASE 1.5, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000, OPTIMA CT660 GT2000 PHASE 1.5, OPTIMA CT660 LITE, OPTIMA CT660 LITE HN P2.5, OPTIMA CT660 MID P1.5 HINO, OPTIMA CT660 MID P20, OPTIMA CT660 MID P2000, OPTIMA CT660 PRO, OPTIMA CT660 PRO GT 2000, OPTIMA CT660 PRO GT1700, OPTIMA CT660 PRO GT1700 ASIR, OPTIMA CT660 PRO GT1700 P20, OPTIMA CT660 PRO GT1700 PHASE, OPTIMA CT660 PRO GT1700 PHASE 1.5, OPTIMA CT660 PRO GT1700V, OPTIMA CT660 PRO GT1700V ASIR, OPTIMA CT660 PRO GT2000, OPTIMA CT660 PRO GT2000 PHASE, OPTIMA CT660 PRO GT2000 PHASE 1.5, OPTIMA CT660 PRO LNG, OPTIMA CT660 PRO LNG P1.5, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LNG P1.5 HINO, OPTIMA CT660 PRO LONG BJG P2.0, OPTIMA CT660 PRO LONG P2.0, OPTIMA CT660 PRO MID P1.5, OPTIMA CT660 PRO MID P1.5 H, OPTIMA CT660 PRO MID P1.5 HINO, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700 BJG, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 BJG2.5, OPTIMA CT660 SE GT 1700 HN2, OPTIMA CT660 SE GT 2000, OPTIMA CT660 SE GT 2000 BJG, OPTIMA CT660 SE GT 2000 BJG2., OPTIMA CT660 SE GT 2000 BJG2.5, OPTIMA CT660 SE GT 2000 HN2, OPTIMA CT660 SE GT 2000 HN2.5, OPTIMA CT660 SE GT2000 MKE2.5, OPTIMA CT660 SE WITH ASIR, OPTIMA CT660 W/PRO PKG-BR, OPTIMA CT660 W/PRO PKG-LA, OPTIMA CT660-64 USA, OPTIMA CT660S, OPTIMA CT660S, OPTIMA CT660S 32 SLICE GT1700, OPTIMA CT660S 32 SLICE HINO, OPTIMA CT660S GT, OPTIMA CT660S GT 1700 HN2.5, OPTIMA CT660S GT 2000, OPTIMA CT660S GT 2000 BJG2.5, OPTIMA CT660S GT 2000 HN2.5, OPTIMA CT660S GT1700 BJG P2.5, OPTIMA CT660S GT1700 HN P2.5, OPTIMA CT660S PRO 32 HINO, OPTIMA CT660S PRO 32 SLICE, OPTIMA CT660S PRO 32 SLICE GT2, OPTIMA CT660S PRO P1.5 HINO, OPTIMACT660_A_G, OPTIMACT660_A_L, OPTIMACT660_N_G, OPTIMACT660_N_L, REVOLUTION EVO, REVOLUTION EVO WITH GT2000 TABLE (H
GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative. GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification.
Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN.
4,007 devices total (products 1, 2, 3)