FDA Recall Terminated

GE Healthcare Revolution EVO, Optima CT660, Optima CT670, Optima CT680. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold under the following product names: CT GOLDSEAL LIGHTSPEED ULTRA, CT OPTIMA 660, CT OPTIMA 660 M40 3.5 LONG BJG, CT OPTIMA 660 M40 GT1700, CT OPTIMA 660 M40 GT1700 BEIJI, CT OPTIMA 660 M40 GT1700 HINO, CT OPTIMA 660 M40 GT2000 BEIJI, CT OPTIMA 660 M40 GT2000 HINO, CT OPTIMA CT 660, CT OPTIMA CT660, CT OPTIMA CT660 M4, CT OPTIMA CT660 M40, CT REVOLUTION EVO 3.5 LONG HIN, CT REVOLUTION EVO VT2000 HINO, CT660 MAR UPGRADE, MOBILE OPTIMA CT 660, OPTI660_A_G_2.0, OPTI660_M40_G, OPTI660_M40_L, OPTIMA 660, OPTIMA 660 128 FOR TW- DIRECT, OPTIMA 660 GT2000 M40 HINO MULTISLICE, OPTIMA 660 M40, OPTIMA 660 M40, OPTIMA 660 M40 3.5 HVY BJG, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 3.5 LONG HINO, OPTIMA 660 M40 3.5 MID, OPTIMA 660 M40 3.5 MID BJG, OPTIMA 660 M40 3.5 MID HINO, OPTIMA 660 M40 3.5/6, OPTIMA 660 M40 3.5/6 HVY BJG, OPTIMA 660 M40 3.5/6 HVY HINO, OPTIMA 660 M40 3.5/6 HVY WAUK, OPTIMA 660 M40 3.5/6 LONG, OPTIMA 660 M40 3.5/6 LONG B, OPTIMA 660 M40 3.5/6 LONG BJG, OPTIMA 660 M40 3.5/6 LONG HINO, OPTIMA 660 M40 3.5/6 LONG WAUK, OPTIMA 660 M40 3.5/6 MID BJ, OPTIMA 660 M40 3.5/6 MID BJG, OPTIMA 660 M40 3.5/6 MID HI, OPTIMA 660 M40 3.5/6 MID HINO, OPTIMA 660 M40 3.5/6 MID WA, OPTIMA 660 M40 3.5/6 MID WAUK, OPTIMA 660 M40 GT1700, OPTIMA 660 M40 GT1700 BEIJING, OPTIMA 660 M40 GT1700 HINO, OPTIMA 660 M40 GT1700 WAUK, OPTIMA 660 M40 GT2000 B, OPTIMA 660 M40 GT2000 BEIJI, OPTIMA 660 M40 GT2000 BEIJING, OPTIMA 660 M40 GT2000 HINO, OPTIMA 660 M40 GT2000 WAUK, OPTIMA 660 M40 MOBILE BEIJING, OPTIMA CT 660, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR 128 M RU, OPTIMA CT660 FREEDOM MD HN2, OPTIMA CT660 M40 GT1700, OPTIMA CT660 M40 GT2000, OPTIMA CT660 M40 GT2000 BEIJIN, OPTIMA CT660 M40 GT2000 HINO, OPTIMA CT660 P2.5, OPTIMA CT660 P25 UPG KIT, OPTIMA CT660 PRO MID, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700, OPTIMA CT680, OPTIMA CT680 KL64 GT1700 BHW, REVOLUTION EVO, REVOLUTION EVO 3.5 HVY BJG, REVOLUTION EVO 3.5 HVY HINO, REVOLUTION EVO 3.5 HVY WAUK, REVOLUTION EVO 3.5 LONG BJG, REVOLUTION EVO 3.5 LONG HINO, REVOLUTION EVO 3.5 LONG WAUK, REVOLUTION EVO 3.5 MID BJG, REVOLUTION EVO 3.5 MID HINO, REVOLUTION EVO 3.5 MID WAUK, REVOLUTION EVO 3.5MID HINO CT, REVOLUTION EVO 3.6 CHN HINO, REVOLUTION EVO 3.6 HVY BJG, REVOLUTION EVO 3.6 HVY HINO, REVOLUTION EVO 3.6 HVY WAUK, REVOLUTION EVO 3.6 LONG, REVOLUTION EVO 3.6 LONG BJG, REVOLUTION EVO 3.6 LONG HIN, REVOLUTION EVO 3.6 LONG HINO, REVOLUTION EVO 3.6 LONG WAUK, REVOLUTION EVO 3.6 MID BJG, REVOLUTION EVO 3.6 MID HINO, REVOLUTION EVO 3.6 MID WAUK, REVOLUTION EVO 3.6B, REVOLUTION EVO 3.6B HVY HINO, REVOLUTION EVO 3.6B LONG, REVOLUTION EVO 3.6B LONG BJG, REVOLUTION EVO 3.6B LONG HINO, REVOLUTION EVO 3.6B MID BJG, REVOLUTION EVO 3.6B MID HINO, REVOLUTION EVO GT1700, REVOLUTION EVO GT2000, REVOLUTION EVO GT2000 HINO, REVOLUTION EVO GT2000X (HINO), REVOLUTION EVO GT2000X HINO, REVOLUTION EVO VT1700 BEIJING, REVOLUTION EVO VT1700 HINO, REVOLUTION EVO VT1700 WAUK, REVOLUTION EVO VT2000 BEIJING, REVOLUTION EVO VT2000 HINO, REVOLUTION EVO VT2000 WAUK, REVOLUTION EVO VT2000X BEIJING, REVOLUTION EVO VT2000X HINO, REVOLUTION EVO VT2000X WAUK, REVOLUTION EVO WITH GT1700V TA, REVOLUTION EVO WITH GT2000 TAB

Recall: Z-2870-2018 · Initiated April 27, 2018

Recall

Recall Number
Z-2870-2018
Event Number
80563
Firm
GE Healthcare Japan Corporation Ge Herusukea.Japan Kabu 4 Chome 7-127 Asahigaoka Hino Japan
FEI Number
3002808529
Product Code
JAK
Status
Terminated
Root Cause
Employee error
Initiated
April 27, 2018
Terminated
August 21, 2020

Description

GE Healthcare Revolution EVO, Optima CT660, Optima CT670, Optima CT680. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. Sold under the following product names: CT GOLDSEAL LIGHTSPEED ULTRA, CT OPTIMA 660, CT OPTIMA 660 M40 3.5 LONG BJG, CT OPTIMA 660 M40 GT1700, CT OPTIMA 660 M40 GT1700 BEIJI, CT OPTIMA 660 M40 GT1700 HINO, CT OPTIMA 660 M40 GT2000 BEIJI, CT OPTIMA 660 M40 GT2000 HINO, CT OPTIMA CT 660, CT OPTIMA CT660, CT OPTIMA CT660 M4, CT OPTIMA CT660 M40, CT REVOLUTION EVO 3.5 LONG HIN, CT REVOLUTION EVO VT2000 HINO, CT660 MAR UPGRADE, MOBILE OPTIMA CT 660, OPTI660_A_G_2.0, OPTI660_M40_G, OPTI660_M40_L, OPTIMA 660, OPTIMA 660 128 FOR TW- DIRECT, OPTIMA 660 GT2000 M40 HINO MULTISLICE, OPTIMA 660 M40, OPTIMA 660 M40, OPTIMA 660 M40 3.5 HVY BJG, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 3.5 LONG BJG, OPTIMA 660 M40 3.5 LONG HINO, OPTIMA 660 M40 3.5 MID, OPTIMA 660 M40 3.5 MID BJG, OPTIMA 660 M40 3.5 MID HINO, OPTIMA 660 M40 3.5/6, OPTIMA 660 M40 3.5/6 HVY BJG, OPTIMA 660 M40 3.5/6 HVY HINO, OPTIMA 660 M40 3.5/6 HVY WAUK, OPTIMA 660 M40 3.5/6 LONG, OPTIMA 660 M40 3.5/6 LONG B, OPTIMA 660 M40 3.5/6 LONG BJG, OPTIMA 660 M40 3.5/6 LONG HINO, OPTIMA 660 M40 3.5/6 LONG WAUK, OPTIMA 660 M40 3.5/6 MID BJ, OPTIMA 660 M40 3.5/6 MID BJG, OPTIMA 660 M40 3.5/6 MID HI, OPTIMA 660 M40 3.5/6 MID HINO, OPTIMA 660 M40 3.5/6 MID WA, OPTIMA 660 M40 3.5/6 MID WAUK, OPTIMA 660 M40 GT1700, OPTIMA 660 M40 GT1700 BEIJING, OPTIMA 660 M40 GT1700 HINO, OPTIMA 660 M40 GT1700 WAUK, OPTIMA 660 M40 GT2000 B, OPTIMA 660 M40 GT2000 BEIJI, OPTIMA 660 M40 GT2000 BEIJING, OPTIMA 660 M40 GT2000 HINO, OPTIMA 660 M40 GT2000 WAUK, OPTIMA 660 M40 MOBILE BEIJING, OPTIMA CT 660, OPTIMA CT660 ASIR, OPTIMA CT660 ASIR 128 M RU, OPTIMA CT660 FREEDOM MD HN2, OPTIMA CT660 M40 GT1700, OPTIMA CT660 M40 GT2000, OPTIMA CT660 M40 GT2000 BEIJIN, OPTIMA CT660 M40 GT2000 HINO, OPTIMA CT660 P2.5, OPTIMA CT660 P25 UPG KIT, OPTIMA CT660 PRO MID, OPTIMA CT660 SE, OPTIMA CT660 SE GT 1700, OPTIMA CT680, OPTIMA CT680 KL64 GT1700 BHW, REVOLUTION EVO, REVOLUTION EVO 3.5 HVY BJG, REVOLUTION EVO 3.5 HVY HINO, REVOLUTION EVO 3.5 HVY WAUK, REVOLUTION EVO 3.5 LONG BJG, REVOLUTION EVO 3.5 LONG HINO, REVOLUTION EVO 3.5 LONG WAUK, REVOLUTION EVO 3.5 MID BJG, REVOLUTION EVO 3.5 MID HINO, REVOLUTION EVO 3.5 MID WAUK, REVOLUTION EVO 3.5MID HINO CT, REVOLUTION EVO 3.6 CHN HINO, REVOLUTION EVO 3.6 HVY BJG, REVOLUTION EVO 3.6 HVY HINO, REVOLUTION EVO 3.6 HVY WAUK, REVOLUTION EVO 3.6 LONG, REVOLUTION EVO 3.6 LONG BJG, REVOLUTION EVO 3.6 LONG HIN, REVOLUTION EVO 3.6 LONG HINO, REVOLUTION EVO 3.6 LONG WAUK, REVOLUTION EVO 3.6 MID BJG, REVOLUTION EVO 3.6 MID HINO, REVOLUTION EVO 3.6 MID WAUK, REVOLUTION EVO 3.6B, REVOLUTION EVO 3.6B HVY HINO, REVOLUTION EVO 3.6B LONG, REVOLUTION EVO 3.6B LONG BJG, REVOLUTION EVO 3.6B LONG HINO, REVOLUTION EVO 3.6B MID BJG, REVOLUTION EVO 3.6B MID HINO, REVOLUTION EVO GT1700, REVOLUTION EVO GT2000, REVOLUTION EVO GT2000 HINO, REVOLUTION EVO GT2000X (HINO), REVOLUTION EVO GT2000X HINO, REVOLUTION EVO VT1700 BEIJING, REVOLUTION EVO VT1700 HINO, REVOLUTION EVO VT1700 WAUK, REVOLUTION EVO VT2000 BEIJING, REVOLUTION EVO VT2000 HINO, REVOLUTION EVO VT2000 WAUK, REVOLUTION EVO VT2000X BEIJING, REVOLUTION EVO VT2000X HINO, REVOLUTION EVO VT2000X WAUK, REVOLUTION EVO WITH GT1700V TA, REVOLUTION EVO WITH GT2000 TAB

Reason

GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.

Action

The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative. GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification.

Distribution

Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN.

Quantity

4,007 devices total (products 1, 2, 3)