FDA Recall Terminated

ALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 15

Recall: Z-2869-2020 · Initiated July 22, 2020

Recall

Recall Number
Z-2869-2020
Event Number
86164
Firm
Globus Medical, Inc.
FEI Number
3004142400
Product Code
MAX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 22, 2020
Posted
August 22, 2020
Terminated
August 19, 2021
Address
2560 General Armistead Ave, Audubon, PA, 19403-5214

Description

ALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 15

Reason

Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

Action

A customer letter dated July 22, 2020 was mailed to customers. In this letter customers are instructed to immediately examine their inventory and determine if they have any of the affected implants. Alternatively, they may return all ALTERA inventory and Globus will sort out the affected implants and replace inventory with conforming implants. Report any adverse effects believed to be associated with the product to Globus Medical immediately by filling out the attached Response Card.

Distribution

Worldwide Distribution