FDA Recall
Terminated
ALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 8
Recall: Z-2866-2020
·
Initiated July 22, 2020
Recall
- Recall Number
- Z-2866-2020
- Event Number
- 86164
- Firm
- Globus Medical, Inc.
- FEI Number
- 3004142400
- Product Code
- MAX
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 22, 2020
- Posted
- August 22, 2020
- Terminated
- August 19, 2021
- Address
- 2560 General Armistead Ave, Audubon, PA, 19403-5214
Description
ALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 8
Reason
Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Action
A customer letter dated July 22, 2020 was mailed to customers. In this letter customers are instructed to immediately examine their inventory and determine if they have any of the affected implants. Alternatively, they may return all ALTERA inventory and Globus will sort out the affected implants and replace inventory with conforming implants. Report any adverse effects believed to be associated with the product to Globus Medical immediately by filling out the attached Response Card.
Distribution
Worldwide Distribution