FDA Recall Terminated

OptiLITE(TM) Products for Laser Surgery, Catalog Numbers: HTM1020F, HTM1040F, HTM1060F, HTM1010F; Manufactured by CONVERGENT LASTER TECHNOLOGIES The Odyssey 30B is approved for numerous medical procedures. Generally, the primary use of the laser (which includes the fibers) is related to urology, but ultimately it is up to the surgeon except in cases where it is contraindicated. The fibers are usually used in cases involving stones anywhere from the bladder up to the kidneys. Occasionally the laser system is used in soft tissue cases (non-cancerous tumors) in and around the urinary system.

Recall: Z-2866-2017 · Initiated June 23, 2017

Recall

Recall Number
Z-2866-2017
Event Number
77653
Firm
Xintec Corp
FEI Number
2938889
Product Code
GEX
Status
Terminated
Root Cause
Labeling False and Misleading
Initiated
June 23, 2017
Terminated
June 20, 2018
Address
1660 S Loop Rd, Alameda, CA, 94502-7091

Description

OptiLITE(TM) Products for Laser Surgery, Catalog Numbers: HTM1020F, HTM1040F, HTM1060F, HTM1010F; Manufactured by CONVERGENT LASTER TECHNOLOGIES The Odyssey 30B is approved for numerous medical procedures. Generally, the primary use of the laser (which includes the fibers) is related to urology, but ultimately it is up to the surgeon except in cases where it is contraindicated. The fibers are usually used in cases involving stones anywhere from the bladder up to the kidneys. Occasionally the laser system is used in soft tissue cases (non-cancerous tumors) in and around the urinary system.

Reason

It has been determined that the reprocessing instructions may not provide sufficiently detailed information for the cleaning, disinfection and sterilization of the products listed below. The preliminary investigation indicates that validation data related to the reprocessing of these devices may not meet newly enacted standards.

Action

The firm, Convergent Laser Technologies, sent an "URGENT - MEDICAL DEVICE RECALL" letter dated 06/23/2017 to its Consignees. The letter described the product, problem and actions to be taken. The consignees were instructed as follows: "Please immediately review your inventory and quarantine any affected product. Please return any affected products within 30 days according to the instructions provided in the attached Acknowledgement and Receipt Form. Upon receipt of the product and response form, we will arrange for a credit to your account. Also, whether or not you are returning any affected product, please send a copy of the Acknowledgement and Receipt Form to our Regulatory Department via fax at 510-832-1600 or via e-mail to [email protected]." If you have any further questions or concerns, please contact our Customer Service Department at 510-832-2130 or 800-848-8200.

Distribution

Worldwide distribution- US (Nationwide) and countries of: Canada and Sri Lanka.

Quantity

4343 units