FDA Recall Terminated

Prime 5th Wheel Stryker Medical Stretchers Model 1105 Intended to transport patients in a horizontal position.

Recall: Z-2865-2011 · Initiated July 19, 2011

Recall

Recall Number
Z-2865-2011
Event Number
59201
Firm
Stryker Medical Division of Stryker Corporation
FEI Number
1831750
Product Code
FPO
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 19, 2011
Posted
July 22, 2011
Terminated
August 29, 2013
Address
3800 East Centre Ave., Portage, MI, 49002

Description

Prime 5th Wheel Stryker Medical Stretchers Model 1105 Intended to transport patients in a horizontal position.

Reason

The firm has determined that the brake rod assemblies of various stretchers that were built with a 5th wheel were manufactured with roll pins that did not meet the materials specifications. Some roll pins may have fractured during the assembly process and could fall out of the brake rod assemblies, rendering the brake or steer systems inoperable from certain pedals or causing a reduction in overa

Action

Stryker Medical sent an "Urgent Medical Device Correction" dated July 19 ,2011 to all affected customers. The letter included; the name and model numbers of the stretchers, the reason for the correction and dates of manufacture. It contained the instructions to check the braking system and if found to be inoperable, to remove it from service, and if operable to check prior to each use. Also, information regarding contacting the recipients of any loaned or sold stretchers to relay the information in the letter and Medwatch reporting information. A self addressed postcard to confirm receipt of the letter was also included. For additional information call (269) 389-6604.

Distribution

Worldwide Distribution: Nationwide (USA) including the states of NY, MI, LA and CA ,and the countries of Canada, China, Spain, Japan, Korea, United Kingdom, South Africa.

Quantity

80