FDA Recall Terminated

DigitalDiagnost Single Detector (Stitching Patient Support) 712020

Recall: Z-2863-2018 · Initiated June 27, 2018

Recall

Recall Number
Z-2863-2018
Event Number
80577
Firm
Philips Medical Systems Gmbh, DMC Development And Manufacturing Ctr. Rontgenstr. 24 Hamburg Germany
FEI Number
3003768251
Product Code
MQB
Status
Terminated
Root Cause
Device Design
Initiated
June 27, 2018
Posted
August 17, 2018
Terminated
April 30, 2024

Description

DigitalDiagnost Single Detector (Stitching Patient Support) 712020

Reason

The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.

Action

On June 27, 2018, Philips Medical Systems issued URGENT FIELD SAFETY NOTICES, Medical Device Recall letters to customers. Customers were encouraged to always check the right position of the hook and ensure that the hook is locked according tot he IFU. Press hook down until it is locked. Make sure that the hook is in position. A Philips Service Engineer will contact impacted customers as soon as the Field Action Kit is ready to be implemented. Should you need to communicate with Philips in regards to this program, please reference FCO-71200185. If you need further information or support concerning this issue, please contact your local Philips representative.

Distribution

US Nationwide; International to 66 countries