Global Focus Marketing and Distribution, Silencer S2200 Centrifuge

Recall: Z-2857-2011 · Initiated July 15, 2011

Recall

Recall Number
Z-2857-2011
Event Number
59288
Firm
Global Focus Marketing & Develop
FEI Number
1000599190
Product Code
JQC
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 15, 2011
Posted
July 27, 2011
Terminated
November 6, 2013
Address
2280 Springlake Rd, Ste 106, Dallas, TX, 75234-5830

Description

Global Focus Marketing and Distribution, Silencer S2200 Centrifuge

Reason

Potential for bucket and insert not seated on the rotor to cause the lid to break and buckets to be ejected.

Action

Global Focus Marketing and Distribution, Ltd issued a press release on 07/19/11 and a URGENT: DEVICE RECALL letter was mailed to the firm's customers on 07/15/11 and 07/21/11. Discontinue use. For questions, contact: (800) 323-4306 or [email protected]

Distribution

Worldwide to USA, United Kingdom, Ireland, Canada, and China.

Quantity

859 units