FDA Recall Terminated

ENCORE(R) UNDERGLOVE, Latex, Powder-Free, Smooth Surgical Gloves, REF 2018460, Size 6, Sterile, Manufactured for Ansell Healthcare, 1635 Industrial Road, Dothan, AL 36303, Made in Sri Lanka Surgical gloves

Recall: Z-2841-2011 · Initiated June 10, 2011

Recall

Recall Number
Z-2841-2011
Event Number
59155
Firm
Ansell Healthcare Products LLC
FEI Number
1019632
Product Code
NPD
Status
Terminated
Root Cause
Other
Initiated
June 10, 2011
Posted
July 18, 2011
Terminated
January 10, 2013
Address
1635 Industrial Rd, Dothan, AL, 36303-5750

Description

ENCORE(R) UNDERGLOVE, Latex, Powder-Free, Smooth Surgical Gloves, REF 2018460, Size 6, Sterile, Manufactured for Ansell Healthcare, 1635 Industrial Road, Dothan, AL 36303, Made in Sri Lanka Surgical gloves

Reason

Gloves do not meet the tensile requirements of ASTM D 3577 reducing the shelf life of the product from 5 years to 2 years.

Action

Ansell Healthcare Products LLC sent an Urgent Voluntary Recall letter dated June 9, 2011, to all affected customers on June 10 and 13, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconstinue use and distribution of any of the affected product and return them to the firm. If the customer distributed gloves further, customers were advised to forward a copy of the notification letter along with their tracking number to their customers. Customers were instructed to complete and return the attached Response Form to Ansell by fax at 1-800-722-8155. Returned product would be replaced or credited to the customer.

Distribution

Nationwide Distribution

Quantity

45692 pairs