ASP-1020 LipoTower with Aspiration Pump. Medical use.

Recall: Z-2821-2015 · Initiated June 12, 2015

Recall

Recall Number
Z-2821-2015
Event Number
72106
Firm
MicroAire Surgical Instruments, LLC
FEI Number
2020601
Product Code
MUU
Status
Terminated
Root Cause
Device Design
Initiated
June 12, 2015
Posted
September 30, 2015
Terminated
May 23, 2016
Address
3590 Grand Forks Blvd, Charlottesville, VA, 22911-9006

Description

ASP-1020 LipoTower with Aspiration Pump. Medical use.

Reason

There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.

Action

The firm has issued FDA recall letters to all consignees. The letter have been sent out by UPS to expedite the delivery.

Distribution

Nationwide Distribution.

Quantity

3 LipoTower with Aspiration Pump