FDA Recall Terminated

Phillips Module Charger, Product #: 989803191031 - Product Usage: This Module Charger is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

Recall: Z-2812-2020 · Initiated June 18, 2020

Recall

Recall Number
Z-2812-2020
Event Number
86025
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
MWI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
June 18, 2020
Terminated
January 10, 2023
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Phillips Module Charger, Product #: 989803191031 - Product Usage: This Module Charger is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

Reason

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Action

Philips Healthcare sent a customer notification letter dated June 2020 explaining the potential issues observed with their products. This letter instructs the consumer to contact the company with any reports of issues associated with this failure.

Distribution

World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.

Quantity

9,705 units