FDA Recall Terminated

InflammaDry(R) AN RPS DIAGNOSTIC SOLUTION, Sample Collectors InflammaDry(R) is a rapid, immunoassay test for the visual, qualitative, in vitro detection of elevated levels of the MMP-9 protein in human tears, from patients suspected of having dry eye. InflammaDry is to be used to aid in the diagnosis of dry eye, in conjunction with other methods of clinical evaluation. This test is intended for prescription use at point-of-care sites.

Recall: Z-2803-2017 · Initiated June 22, 2017

Recall

Recall Number
Z-2803-2017
Event Number
77640
Firm
Quidel Corporation
FEI Number
2024674
Product Code
PFQ
Status
Terminated
Root Cause
Process design
Initiated
June 22, 2017
Terminated
June 7, 2019
Address
10165 McKellar Ct, San Diego, CA, 92121-4201

Description

InflammaDry(R) AN RPS DIAGNOSTIC SOLUTION, Sample Collectors InflammaDry(R) is a rapid, immunoassay test for the visual, qualitative, in vitro detection of elevated levels of the MMP-9 protein in human tears, from patients suspected of having dry eye. InflammaDry is to be used to aid in the diagnosis of dry eye, in conjunction with other methods of clinical evaluation. This test is intended for prescription use at point-of-care sites.

Reason

In-house testing revealed the presence of plastic spurs at the tips of some Sample Collectors.

Action

Quidel sent an Urgent Medical Device Recall letter dated June 22, 2017, to all affected customers. The notice instructed consignees to check inventory for affected lots (11705359 and 1705405), to remove and destroy any affected sample collectors from affected kits, and complete and return the attached Inventory Assessment/Certificate of Destruction form to Quidel by fax at 858-552-7905 or email to [email protected]. Any questions by consignees should be directed to Quidel's Technical Support Team at 800-874-1517 or 858-552-1100, or by email at [email protected]. It is requested that consignees keep recall notification and response in their records.

Distribution

Nationwide Distribution including AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TX, VA, WA, & WV.

Quantity

1078 kits, 21,560 sample collectors