FDA Recall Terminated

Screw Removal Case Assembly, Part Number 1985BASE - Product Usage: The Wright Medical hardware removal kit is designed to facilitate removal of both plates and screws used in foot and ankle procedures. Along with a complete range of drivers, the kit also features driver handles and a range of k-wires to assist in removal of cannulated screws. This product is a multiple use, non-sterile instrument.

Recall: Z-2797-2020 · Initiated July 28, 2020

Recall

Recall Number
Z-2797-2020
Event Number
85982
Firm
Wright Medical Technology, Inc.
FEI Number
3010667733
Product Code
LRO
Status
Terminated
Root Cause
Device Design
Initiated
July 28, 2020
Terminated
November 16, 2022
Address
1023 Cherry Rd, Memphis, TN, 38117-5423

Description

Screw Removal Case Assembly, Part Number 1985BASE - Product Usage: The Wright Medical hardware removal kit is designed to facilitate removal of both plates and screws used in foot and ankle procedures. Along with a complete range of drivers, the kit also features driver handles and a range of k-wires to assist in removal of cannulated screws. This product is a multiple use, non-sterile instrument.

Reason

Hardware removal kit does not include drivers to remove Dart-Fire 2.5 and 3.0mm Headless screws.

Action

Wright disseminated the notices by Fed Ex and email on 07/28/2020. The notices requested the following actions: " Cease using the screw and driver combinations: Dart-Fire 2.5mm Headless Screw with HCS-056-25 1.5 Hex Driver and Dart-Fire 3.0mm Headless Screw and 44112001 2.0 Hex Driver " Disseminate this notice to all those who need to be aware within your organization. " Ensure that you have the right products for removal surgery through pre-operative planning and coordinating with your Wright Medical Representative. The firm will provide a new kit when available.

Distribution

US Nationwide distribution including in the states of CO, TX, OR, KX, SC, CA, AR, NJ, UT, VA, MO, NC, FL, IL.

Quantity

9 units