FDA Recall Terminated

Sysmex Stromatolyser-IM; a lytic reagent for use with the XE-2100, XE-2100L and XE-5000 Series Sysmex Automated Hematology Analyzers; the reagent solution is supplied in a 10-liter cubetainer (cardboard cube with a 10-liter plastic liner); Sysmex America, Inc., Mundelein, IL 60060; catalog number SIM-220A. The product is a lytic reagent used to lyse (destroy) red blood cells for hemoglobin determinations or aid in the counting of white blood cells.

Recall: Z-2797-2011 · Initiated June 23, 2009

Recall

Recall Number
Z-2797-2011
Event Number
58916
Firm
Sysmex America, Inc.
FEI Number
1000515253
Product Code
GGK
Status
Terminated
Root Cause
Environmental control
Initiated
June 23, 2009
Posted
July 13, 2011
Terminated
July 13, 2011
Address
1 Nelson C White Pkwy, Mundelein, IL, 60060-9528

Description

Sysmex Stromatolyser-IM; a lytic reagent for use with the XE-2100, XE-2100L and XE-5000 Series Sysmex Automated Hematology Analyzers; the reagent solution is supplied in a 10-liter cubetainer (cardboard cube with a 10-liter plastic liner); Sysmex America, Inc., Mundelein, IL 60060; catalog number SIM-220A. The product is a lytic reagent used to lyse (destroy) red blood cells for hemoglobin determinations or aid in the counting of white blood cells.

Reason

The affected lots of Stromatolyser-IM lysing reagent was found to be contaminated with Pseudomonas bacteria, which resulted in rising IMI channel background counts, increased false "Platelet Clump" flags and a strong sulfur-like odor when used on the XE-2100, XE-2100L and XE-5000 analyzers.

Action

Sysmex America, Inc. sent a "MEDICAL DEVICE RECALL" letter dated June 2009 to all affected customers. The letter identifies the product, problem, and actions to be taken. The letter instructs customers to discontinue use of the affected SIM lots, and to remove affected lots from their analyzers as soon as possible to avoid long-term performance problems on the analyzer. A Stromatolyser -IM (220-A) Voluntary Recall form was attached for customers to complete and return via fax at 800-218-2711. The form will be used to place orders to replace the affected product. Any questions concerning the information contained in the recall letter are directed to the Sysmex Technical Assistance Center at 1-888-879-7639. The firm sent a second notice to non-responding accounts on September 14, 2009. A copy of the June 26, 2009 recall letter and the fax back form were included with the letter. The customers were informed that Sysmex had not received their completed fax back form and were requested to respond to confirm the notification was received.

Distribution

Worldwide Distribution--USA (nationwide) and the countries of Brazil, Canada and Honduras.

Quantity

4,005 cubetainers