Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.
Recall
- Recall Number
- Z-2795-2018
- Event Number
- 80467
- Firm
- Hamilton Medical AG Via Crusch 8 Bonaduz Switzerland
- FEI Number
- 3001421318
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 9, 2018
- Terminated
- March 18, 2019
Description
Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pressure ventilation support to adults and pediatrics. The device is used in the in the intensive care ward, in the recovery room, and during transfer of ventilated patients within the hospital.
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.
Hamilton Medical notified customers on about 03/09/2018 via letter. Instructions included that a Hamilton Medical Field Service Representative will be in contact with each customer to schedule the correction. Questions or concerns can be directed to Hamilton Technical Support at 1-800-426-6331, option #2.
US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.
16 devices