Guidant RAPIDO Cut-Away Guiding Catheter, 6F. Part Numbers: 66676-201 and 667552-100. For use in cardiac diagnostic and therapeutic procedures.

Recall: Z-2793-2011 · Initiated January 7, 2011

Recall

Recall Number
Z-2793-2011
Event Number
58979
Firm
Abbott Vascular
FEI Number
2024168
Product Code
DQY
Status
Terminated
Root Cause
Process control
Initiated
January 7, 2011
Posted
July 12, 2011
Terminated
June 26, 2012
Address
26531 Ynez Rd, Temecula, CA, 92591-4630

Description

Guidant RAPIDO Cut-Away Guiding Catheter, 6F. Part Numbers: 66676-201 and 667552-100. For use in cardiac diagnostic and therapeutic procedures.

Reason

Five lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.

Action

Abbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product

Distribution

Nationwide distribution, and Ireland.

Quantity

389 units total