MegaPower Electrosurgical Generator, Product code 1000, GTIN 10614559103715
Recall
- Recall Number
- Z-2792-2018
- Event Number
- 80327
- Firm
- Megadyne Medical Products, Inc.
- FEI Number
- 1721194
- Product Code
- GEI
- Status
- Terminated
- Root Cause
- Use error
- Initiated
- May 15, 2018
- Posted
- August 13, 2018
- Terminated
- January 30, 2024
- Address
- 11506 S State St, Draper, UT, 84020-9453
Description
MegaPower Electrosurgical Generator, Product code 1000, GTIN 10614559103715
If two devices are plugged into the same channel/port in the generator, both devices activate. If one device is left laying on the patient, the result could be burning of the patient.
The firm sent letters in June 2018, informing their customers of the issue. The firm will send labels that can be applied to the product to remind them not to plug 2 active devices into the same channel at a time. Updated 08/13/19 Starting on 06/29/18, updated Medical Device Correction Notifications and labels were mailed via second day service. Customers were asked to complete and return Business Reply Forms.
Worldwide distribution. US Nationwide, Mexico, Europe, Middle East, Latin America, Africa, and Asia Pacific Markets
8380 units