Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound imaging or fluid flow analysis of the human body.
Recall
- Recall Number
- Z-2782-2011
- Event Number
- 57012
- Firm
- Bard Access Systems
- FEI Number
- 3006260740
- Product Code
- IYN
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- June 21, 2011
- Posted
- July 11, 2011
- Terminated
- July 11, 2012
- Address
- 605 North 5600, West Salt Lake City, UT, 84116-3738
Description
Site Rite Vision Ultrasound System, REF 9770032, Bard Access Systems, Inc. Intended for diagnostic ultrasound imaging or fluid flow analysis of the human body.
Purchasing erroneously ordered the incorrect fan using the distributor's part number instead of using the specified manufacturer part number.
Bard Access Systems, Inc sent out an "Urgent" Medical Device Recall (User Level)" dated June 2011 to all affected customers The letter included: product name, description of problem and requested the return of units bearing "B50" on the fan. Bard Access Systems, Inc. will replace the incorrect fan with the correct fan. For additional information call 1-724-779-3000.
Nationwide (USA) distribution including the states of AL, AR, AZ, CA, CO, FL, IA, IN, MD, MI, MO, MS, NC, NJ, NM, NY, OH, PA, RI, SC, TX, UT, VA, VT, and WA.
86 units