FDA Recall Terminated

NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols delivered directly to the patient for breathing. Designed to be used in either a hospital or homecare environment.

Recall: Z-2781-2015 · Initiated August 4, 2015

Recall

Recall Number
Z-2781-2015
Event Number
71909
Firm
Salter Labs 2365 Camino Vida Roble
FEI Number
1314417
Product Code
CAF
Status
Terminated
Root Cause
Component change control
Initiated
August 4, 2015
Posted
September 17, 2015
Terminated
May 6, 2016
Address
Carlsbad, CA, 92011-1505

Description

NebuTech Nebulizer, Model number: Series 8660, 8960, 8961, 8966, 8984. Nebulizer used to generate aerosols delivered directly to the patient for breathing. Designed to be used in either a hospital or homecare environment.

Reason

The Inspiratory valve cap on the nebulizer reservoir tower may be loose and may detach before or during treatment.

Action

Salter Labs sent a Medical Device Recall letter dated July 24 2015. The letter identified the affected product, problem and actions to be taken. The letter states that all affected product can be returned to Salter Labs for replacement at no charge with product specifically inspected for a tighter fit on the inspiratory valve cap. Customers were instructed to complete and return the Return Response Form as soon as possible. For any questions contact Salter Labs Customer Support at 800-421-0024 or the local Salter Labs representative.

Distribution

Worldwide Distribution - US Nationwide and in the countries of Greece, UK, Belgium, Australia, and Switerzland.

Quantity

8,975