FDA Recall Terminated

Eclipse Treatment Planning System; Release 6.5 through 10 (build 7.3.10 through 10.0.34) Varian Medical Systems, Inc. Used to plan radiotherapy treatments for patients with Malignant or benign diseases. Plans beam irradiation with photon, electron and proton beams, as well as for internal radiation treatments.

Recall: Z-2780-2011 · Initiated May 13, 2011

Recall

Recall Number
Z-2780-2011
Event Number
58969
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
May 13, 2011
Posted
July 11, 2011
Terminated
August 15, 2014
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Eclipse Treatment Planning System; Release 6.5 through 10 (build 7.3.10 through 10.0.34) Varian Medical Systems, Inc. Used to plan radiotherapy treatments for patients with Malignant or benign diseases. Plans beam irradiation with photon, electron and proton beams, as well as for internal radiation treatments.

Reason

In affected Eclipse versions, user editing of the MLC Leaf positions may not automatically invalidate calculated dose distributions.

Action

Varian Medical Systems, Inc sent a "Urgent Medical Device Correction Urgent Field Safety Notice" letter dated May 13, 2011 to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter included the description of the product, problem and recommendations and includes the release of a service pack in the future. For additional information please contact the Varian Oncology Help Desk at 888-VARIAN5 (888- 827-4265)

Distribution

Worldwide distribution: Nationwide (USA) and the country of Canada.

Quantity

10,097***12-12-12 AMENDED TO: 10664 units***