FDA Recall
Terminated
Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.
Recall: Z-2778-2015
·
Initiated June 30, 2015
Recall
- Recall Number
- Z-2778-2015
- Event Number
- 71695
- Firm
- Fujifilm Medical Systems U.S.A., Inc.
- FEI Number
- 2431293
- Product Code
- FDT
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 30, 2015
- Posted
- September 17, 2015
- Terminated
- May 25, 2017
- Address
- 10 Highpoint Dr, Wayne, NJ, 07470-7431
Description
Fujifilm Medical Endoscope for the duodenum and upper G.I. tract. ED-530XT, ED-250XT5, ED-250XL5, ED-450XT5, and ED-450XL5 Endoscope Operation Manuals.
Reason
The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.
Action
FujiFilm sent a Voluntary Field Correction/Verification Form dated June 30, 2015 to their affected customers.
Distribution
Distributed in the US and Canada.
Quantity
256 units