FDA Recall Terminated

Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120 ml fill. R2: HDPE plastic bottle. 90 ml fill.

Recall: Z-2773-2015 · Initiated December 12, 2014

Recall

Recall Number
Z-2773-2015
Event Number
71974
Firm
Pointe Scientific, Inc.
FEI Number
1827821
Product Code
CGX
Status
Terminated
Root Cause
Process control
Initiated
December 12, 2014
Posted
September 16, 2015
Terminated
October 29, 2015
Address
5449 Research Dr, Canton, MI, 48188-2261

Description

Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120 ml fill. R2: HDPE plastic bottle. 90 ml fill.

Reason

Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.

Action

On 12/12/2014, Field Correction Notice notifications were sent to the affected distributors via electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Any questions concerning the corrective action may be directed to Pointe Scientifics Technical Service Department at 1-800-757-5313.

Distribution

Worldwide distribution. US nationwide, Singapore, Kenya, Bolivia, Mexico, Greece, Guyana, and Philippines.

Quantity

93 kits