FDA Recall Terminated

OmniTom Computed tomography x-ray system, Model NL5000, all NL5000 systems with software version before 05.01.01 - Product Usage: is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck.

Recall: Z-2772-2020 · Initiated July 15, 2020

Recall

Recall Number
Z-2772-2020
Event Number
86052
Firm
NeuroLogica Corporation
FEI Number
3004938766
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
July 15, 2020
Terminated
July 13, 2021
Address
14 Electronics Ave, Danvers, MA, 01923-1011

Description

OmniTom Computed tomography x-ray system, Model NL5000, all NL5000 systems with software version before 05.01.01 - Product Usage: is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck.

Reason

There are potential issues related to the operational state of the scanner, specifically for transport and inadvertent system shutdowns. There could be a possibility of potential delay in patient treatment if proper use instructions are not followed.

Action

Urgent Medical Device Correction notification letters were sent to customers beginning 7/15/20. Users may continue to use the NL5000 system if proper instructions for use are followed as described in user manual (1-NL5000-060). This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. The system label is located near the main input plug or on the side of the system. NeuroLogica will undertake the voluntary corrective action considering the possibility of delay in patient treatment if the scanner is not operational. The hardware and software upgrades provide additional safeguards by improving the design to minimize any risks of the above situations from happening. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction. All corrections in the field will be completed by December 31, 2020. If you need any further information or have any concerns with this issue, then please contact your local representative. Customer Reply Form This reply form is to confirm receipt of the enclosed NeuroLogica Medical Device Field Safety Notice [FSN-NL5000_071620] regarding the potential issues related to the operational state of the scanner and the required software and hardware upgrades. Please read the Notice and indicate the appropriate answers to the questions below.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Canada, India, Portugal, Singapore, Slovak Republic, Thailand, Turkey, UK.

Quantity

35