FDA Recall Terminated

Beckman Coulter - Manual Inhibin A ELISA Reagent Kit; Part Number DSL-10-28100-1 The Inhibin A enzyme-linked immunosorbent assay (ELISA) kit provides materials for the quantitative measurement of dimeric inhibin A in human serum and plasma. This assay is intended for in vitro diagnostic use as an aid in the diagnosis and monitoring of various hormonal reproductive disorders.

Recall: Z-2772-2011 · Initiated March 31, 2011

Recall

Recall Number
Z-2772-2011
Event Number
58720
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
ndr
Status
Terminated
Root Cause
Process design
Initiated
March 31, 2011
Posted
July 11, 2011
Terminated
July 11, 2012
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Beckman Coulter - Manual Inhibin A ELISA Reagent Kit; Part Number DSL-10-28100-1 The Inhibin A enzyme-linked immunosorbent assay (ELISA) kit provides materials for the quantitative measurement of dimeric inhibin A in human serum and plasma. This assay is intended for in vitro diagnostic use as an aid in the diagnosis and monitoring of various hormonal reproductive disorders.

Reason

The recall was initiated because patient results are falsely elevated by approximately 30% because the reference standard of the Manual Inhibin A ELISA assay has shifted from the assays original calibration. Erroneously elevated results may lead to unnecessary ultrasound testing to confirm gestational age, patient psychological stress until diagnostic testing is completed, and/or medical decis

Action

Beckman Coulter sent an "Urgent Product Correction" letter dated March 31, 2011 to all affected customers. The letter included an explanation of the problem and recommendations including; discontinue use of the kit and a Response Form to fill out and return. For additional information contact the Beckman Coulter Customer Support Center at 1-800-854-3633 or your local Beckman Coulter representative.

Distribution

Worldwide distribution: (USA) Nationwide and the countries of Armenia, Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Jordan, Republic of Korea, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Peru, Portugal, Romania, Russian Federation, Serbia, Slovenia, Spain, Thailand, Tunisia, Turkey, United Arab Emirates, and United Kingdom.