VITEK 2 Gram Negative test kits containing colistin (cs01n)
Recall
- Recall Number
- Z-2770-2017
- Event Number
- 77373
- Firm
- Biomerieux Inc
- FEI Number
- 1950204
- Product Code
- PTJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 16, 2017
- Posted
- June 27, 2017
- Terminated
- June 19, 2019
- Address
- 595 Anglum Rd, Hazelwood, MO, 63042-2320
Description
VITEK 2 Gram Negative test kits containing colistin (cs01n)
The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).
Customers were notified on or about 05/16/2017 and then notified again with an updated letter. Instructions to customers included requesting that the attached acknowledgement form be completed, distribute the recall notification to customers, and provide the notice to future customers.
Domestic: NC. Foreign: Argentina, Austria, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, France, Germany, Guatemala, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Netherlands Antilles, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Singapore, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, Venezuela, & Vietnam.
1,520,000 cards