FDA Recall Terminated

VITEK 2 Gram Negative test kits containing colistin (cs01n)

Recall: Z-2770-2017 · Initiated May 16, 2017

Recall

Recall Number
Z-2770-2017
Event Number
77373
Firm
Biomerieux Inc
FEI Number
1950204
Product Code
PTJ
Status
Terminated
Root Cause
Other
Initiated
May 16, 2017
Posted
June 27, 2017
Terminated
June 19, 2019
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

VITEK 2 Gram Negative test kits containing colistin (cs01n)

Reason

The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).

Action

Customers were notified on or about 05/16/2017 and then notified again with an updated letter. Instructions to customers included requesting that the attached acknowledgement form be completed, distribute the recall notification to customers, and provide the notice to future customers.

Distribution

Domestic: NC. Foreign: Argentina, Austria, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, France, Germany, Guatemala, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Netherlands Antilles, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Singapore, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, Venezuela, & Vietnam.

Quantity

1,520,000 cards