Lumenis VersaCut+ Tissue Morcellator GA-0007500 The Lumenis VersaCut / VersaCut+ Tissue Morcellator System is intended for use under direct or endoscopic visualization for the morcellation and removal of dissected tissue during pelviscopic, laparoscopic, percutaneous, and open surgical procedures whenever access to the surgical site is limited.
Recall
- Recall Number
- Z-2770-2015
- Event Number
- 71851
- Firm
- Lumenis 13 Hayetzira St., Yokneam Ind. Park Yokneam Israel
- FEI Number
- 3004135191
- Product Code
- GCJ
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 30, 2015
- Posted
- September 16, 2015
- Terminated
- May 25, 2018
Description
Lumenis VersaCut+ Tissue Morcellator GA-0007500 The Lumenis VersaCut / VersaCut+ Tissue Morcellator System is intended for use under direct or endoscopic visualization for the morcellation and removal of dissected tissue during pelviscopic, laparoscopic, percutaneous, and open surgical procedures whenever access to the surgical site is limited.
There is a probability the hand piece may operate inconsistently and unpredictably during the morcellation procedure resulting in damage to non-target tissue.
Lumenis sent an Urgent Medical Device Recall letter on August 7, 2015, to all consignees. Consignees were instructed to destroy the operator manual, and return the affected device to Lumenis along with the completed Recall Verification form. Consignees with questions were instructed to call 801-656-2663.
Worldwide Distribution - US including CA, PA, MA, FL, and NY, and Internationally to Italy, Germany, Australia, China, Great Britian, United Kingdom, Taiwan, France, Peru, Mexico, Turkey, Spain, and Russia.
54 devices with 108 hand pieces