FDA Recall Terminated

Sterile Drape, SK6000 PAQ drape pouch (for use with the SPY Fluorescent System (SPY 2000/2001)) The SPY Fluorescent Imaging System is intended to provide fluorescent images for the visual assessment of blood flow in vessels and related tissue perfusion during cardiovascular surgical procedures. For use in intraoperative visual assessment of the coronary vasculature and bypass grafts during coronary artery bypass (BABG) surgery.

Recall: Z-2767-2016 · Initiated August 18, 2016

Recall

Recall Number
Z-2767-2016
Event Number
75006
Firm
NOVADAQ TECHNOLOGIES INC.
FEI Number
3012345110
Product Code
IZI
Status
Terminated
Root Cause
Packaging
Initiated
August 18, 2016
Terminated
December 9, 2016
Address
8329 Eastlake Dr, Unit 101, Burnaby Canada

Description

Sterile Drape, SK6000 PAQ drape pouch (for use with the SPY Fluorescent System (SPY 2000/2001)) The SPY Fluorescent Imaging System is intended to provide fluorescent images for the visual assessment of blood flow in vessels and related tissue perfusion during cardiovascular surgical procedures. For use in intraoperative visual assessment of the coronary vasculature and bypass grafts during coronary artery bypass (BABG) surgery.

Reason

Novadaq is informing SK6000 PAQ users that a pinhole through the drape pouch film material was identified in some SK6000 PAQs during shipping validation testing.

Action

The firm, Novadaq, sent an "URGENT DEVICE CORRECTION: SK6000 PAQ Drape Pouch" letter dated 8/18/2016 to affected consignees. The Firm instructed consignee to cease distribution of affected product; accept the shipping-validated replacement PAQ; complete and sign the enclosed "Acknowledgment of Receipt" form, scan it and email to [email protected]. If you have any questions, contact Director of Regulatory Affairs at 1-905-629-3822 ext 205 or email: [email protected]. .

Distribution

US Distribution in states of: FL, MI, and VA.

Quantity

20 units