FDA Recall Terminated

NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.

Recall: Z-2767-2015 · Initiated July 29, 2015

Recall

Recall Number
Z-2767-2015
Event Number
72015
Firm
Nico Corp.
FEI Number
3007456622
Product Code
GEY
Status
Terminated
Root Cause
Process control
Initiated
July 29, 2015
Posted
September 15, 2015
Terminated
September 22, 2015
Address
250 E 96th St, ste 125, Indianapolis, IN, 46240-3781

Description

NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.

Reason

Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.

Action

The firm, Nico Corporation, sent an "URGENT: NICO Console Recall" notification dated 7/31/15 via fax to its customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The customers were instructed to return the console and complete and return the Recall Response Form via fax to: 317-683-0305 or email to: [email protected]. The firm will provide customers with a Myriad Console loaner kit. If you have any questions, contact Director of Quality and Regulatory, at 317-660-7118 ext 104 or email to: [email protected].

Distribution

US Distribution to states of: MD, IN, CA, and NC.

Quantity

6 units