VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The system is intended for use to perform in vitro diagnostic organism identification and antimicrobial susceptibility testing in human applications.
Recall
- Recall Number
- Z-2767-2011
- Event Number
- 58981
- Firm
- Biomerieux Inc
- FEI Number
- 1950204
- Product Code
- LTW
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 2, 2011
- Posted
- July 9, 2011
- Terminated
- June 6, 2013
- Address
- 595 Anglum Rd, Hazelwood, MO, 63042
Description
VITEK 2 System, VITEK 2 60 110V (Catalog #27225) and VITEK 2 60 220V (Catalog #27226). VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The system is intended for use to perform in vitro diagnostic organism identification and antimicrobial susceptibility testing in human applications.
Cards may not load into the incubator correctly causing them to be loaded in the wrong order which could result in the results being linked to the wrong accession number and reporting results erroneously.
Biomerieux, Inc. sent an "URGENT PRODUCT CORRECTION NOTICE" dated June 2, 2011 to all affected customers. The letter describes the products, problem, and actions to be taken by the customers. The letter explains how to recognize the issue and instructs customers to take certain steps to resolve the problem. Additionally, an Acknowledgement Form is attached to the letter for customers to complete and return via fax to 919-314-4539. A firmware update will be applied by Biomerieux, Inc. technical support representatives. Questions regarding this recall are directed to the Customer Service Department at 1-800-682-2666, option 3.
Worldwide Distribution-- USA (nationwide) and the countries of Austria, Australia, Belgium, Brazil, Canada, China, Colombia, Germany, France, India, Israel, Italy, Japan, Korea, Mexico, The Netherlands, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirate, and United Kingdom.
627 instruments