FDA Recall Terminated

Boston Scientific, WATCHMAN Left Atrial Appendage Closure Device with Delivery System, 21mm x 12F ( M635WU21060, WU21060, 24mm x 12F(M635WU24060, WU2406) , 27mm x 12F(M635WU27060, WU2706) , 30mm x 12F(M635WU30060, WU3006) , 33mm x 12F( M635WU33060, WU3306). For cardiac use.

Recall: Z-2765-2015 · Initiated August 25, 2015

Recall

Recall Number
Z-2765-2015
Event Number
72045
Firm
Boston Scientific Corporation
FEI Number
3002095335
Product Code
NGV
Status
Terminated
Root Cause
Use error
Initiated
August 25, 2015
Terminated
April 4, 2017
Address
1 Scimed Pl, Maple Grove, MN, 55311-1565

Description

Boston Scientific, WATCHMAN Left Atrial Appendage Closure Device with Delivery System, 21mm x 12F ( M635WU21060, WU21060, 24mm x 12F(M635WU24060, WU2406) , 27mm x 12F(M635WU27060, WU2706) , 30mm x 12F(M635WU30060, WU3006) , 33mm x 12F( M635WU33060, WU3306). For cardiac use.

Reason

Cross-threading of the hemostasis valve may occur if it is tightened with the dilator in place, possibly preventing subsequent sealing of the valve when desired. This Field Correction reinforces the Directions for Use (DFU) and gives further guidance on correct use of the hemostasis valve to avoid cross-threading and to securely seal the valve, minimizing potential for undesired blood leakage.

Action

Consignees were sent a Boston Scientific "Urgent Field Correction" letter dated August 25, 2015. The letter was addressed to Materials manager / Field Action Contact. The letter described the problem and the product affected. The Field Correction reinforces existing Directions for USE (DFU) and provides further guidance regarding the correct us of the hemostatis valve in order to avoid cross-threading and to securely seal the valve, minimizing the poential for undesirable blood leakage. The consignees were informed of the action required which stated that no product is required to be returned and advised to consignees that they must complete the Customer Acknowledgement Form and return it to Boston Scientific (MapleGroveFieldAction [email protected] or Fax to Field action Center 1-866-213-1806). The letter also provided a section of the Updated Directions for Use which were highlighted in blue. If the consignee is a distributor the Field Correction is to the hospital level. For questions, customers may contact their Local Sales Representative.

Distribution

Worldwide Distribution-US (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Oman, Panama, Philippines, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, White Russia.

Quantity

29,169 units total