Patient Cable. Part Numbers: 4460, 6735. Sterile within Chipboard Boxes within corrugated cartons. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP)
Recall
- Recall Number
- Z-2764-2016
- Event Number
- 74893
- Firm
- Blackrock Microsystems, LLC
- FEI Number
- 3007323246
- Product Code
- GWL
- Status
- Terminated
- Root Cause
- Labeling False and Misleading
- Initiated
- August 10, 2016
- Posted
- September 8, 2016
- Terminated
- August 21, 2017
- Address
- 630 S Komas Dr, Ste 200, Salt Lake City, UT, 84108-1247
Description
Patient Cable. Part Numbers: 4460, 6735. Sterile within Chipboard Boxes within corrugated cartons. Product Usage: The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP)
Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type CF Applied Parts.
Blackrock Microsystems sent an Urgent Medical Device letter dated August 12, 2016 via email. The letter identified the affected product, problem and actions to be taken. Customers were instructed not use the amplifier in application requiring a type CF applied part; perform a physical count of the inventory of amplifiers and patient cables with the part numbers listed. For questions regarding this Correction call 801-582- 5533, ext. 240.
Worldwide Distribution - US Nationwide and Canada,
120 total cables