FDA Recall Terminated

Clinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Recall: Z-2764-2011 · Initiated May 18, 2011

Recall

Recall Number
Z-2764-2011
Event Number
58976
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Other
Initiated
May 18, 2011
Posted
July 8, 2011
Terminated
March 13, 2012
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Clinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Reason

The Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.

Action

Varian Medical Systems, Inc. Oncology Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letters on May 18, 2011, via certified mail, return receipt requested. The letter identified the product, problem and recommended actions to be taken. The letter states that customers will be contacted by Varian Service to arrange for replacement of the recalled product. For additional information contact Varian Medical Systems, Inc at 888- 827-4265.

Distribution

Worldwide Distribution -- USA (nationwide) and countries of: Australia, Columbia, Denmark, France, Ireland, Italy, Japan, Morocco, Poland, Portugal, Saudi Arabia, South Korea, Spain, and United Kingdom.

Quantity

93