FDA Recall Terminated

XGAUZE Trauma Dressing, catalog number FIN-0002-01. XGAUZE is a 3" x 36" z-folded rayon/polyester dressing embedded with individual compressed cellulose sponges. Upon contact with blood or exudate, the sponges expand, thus increasing the dressing's capacity to absorb fluids. XGAUZE is packed in a pre-formed sterile barrier system (pouch sealed on three sides); one side is a PET/foil/PE film, and the other side is a PET/aluminum oxide/PE film. Packages are vacuum sealed and terminally sterilized via gamma radiation.

Recall: Z-2762-2015 · Initiated August 12, 2015

Recall

Recall Number
Z-2762-2015
Event Number
72019
Firm
Revmedx Inc
FEI Number
3009035955
Product Code
NAB
Status
Terminated
Root Cause
Package design/selection
Initiated
August 12, 2015
Posted
September 11, 2015
Terminated
September 17, 2015
Address
25999 Sw Canyon Creek Rd, Wilsonville, OR, 97070-9677

Description

XGAUZE Trauma Dressing, catalog number FIN-0002-01. XGAUZE is a 3" x 36" z-folded rayon/polyester dressing embedded with individual compressed cellulose sponges. Upon contact with blood or exudate, the sponges expand, thus increasing the dressing's capacity to absorb fluids. XGAUZE is packed in a pre-formed sterile barrier system (pouch sealed on three sides); one side is a PET/foil/PE film, and the other side is a PET/aluminum oxide/PE film. Packages are vacuum sealed and terminally sterilized via gamma radiation.

Reason

The XGAUZE Trauma Dressing was recalled because the packages may acquire small tears or pinholes during shipping due to use of shipping containers that lacked adequate protective materials. If present, these small tears or pinholes may disrupt package integrity and render the device unsterile.

Action

RevMedx, Inc. sent the Urgent Medical Device Correction and Removal XGAUZE Trauma Dressing letter to US Customers on August 12, 2015. Customers are instructed to inspect packages for signs of damages or loss of vacuum. The affected packages will appear puffy and the dressing will be loose inside. The non-affected packages will have a good vacuum intact seal and the packaging material should contour snugly against the dressing. Customers are advised to return any packages appear to have damaged or loss of vacuum. Any undamaged packages may be used as indicated. Customers with question can call the firm at (503) 218-2172 and fax the Return Response Form via fax to (503) 218-2274 or send via email to [email protected].

Distribution

US Distribution to the states of : Arizona, Idaho, North Carolina, Tennessee and Virginia.

Quantity

5004 units