Hemocor High Performance Hemoconcentrator, HPH700, sterilized with ethylene oxide The Hemocor HPH700 Hemoconcentrator is intended for use as an ultrafiltration system to remove excess fluid during and/or following cardiopulmonary bypass procedures where acute hemodilution is employed.
Recall
- Recall Number
- Z-2761-2015
- Event Number
- 72122
- Firm
- Medivators, Inc.
- FEI Number
- 2150060
- Product Code
- KDI
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- August 3, 2015
- Posted
- September 11, 2015
- Terminated
- November 4, 2015
- Address
- 14605 28th Ave N, Plymouth, MN, 55447-4822
Description
Hemocor High Performance Hemoconcentrator, HPH700, sterilized with ethylene oxide The Hemocor HPH700 Hemoconcentrator is intended for use as an ultrafiltration system to remove excess fluid during and/or following cardiopulmonary bypass procedures where acute hemodilution is employed.
Hemoconcentrators may exhibit low ultrafiltration performance that is below product specification.
Medivators sent an "Urgent Medical Device" recall letter, dated August 4, 2015, via email. The letter described the issue, identified the affected lot, requested the product be quarantined, and to contact Medivators for return instructions and credit or no-charge replacement product. A response form was requested to be returned to Medivators. Customers with questions can contact Medivators Product at 763-553-3348.
US Distribution to the states of CA and MI.
840 units