Active Cortisol EIA, 96 Wells, DSL-10-2000 The DSL-10-2000 Active Cortisol Enzyme Immunoassay (EIA) kit provides materials for the quantitative measurement of cortisol in serum, plasma or urine. This assay is intended for in vitro diagnostic use.
Recall
- Recall Number
- Z-2761-2011
- Event Number
- 55227
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- CGR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 20, 2009
- Posted
- July 8, 2011
- Terminated
- April 27, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92821-6208
Description
Active Cortisol EIA, 96 Wells, DSL-10-2000 The DSL-10-2000 Active Cortisol Enzyme Immunoassay (EIA) kit provides materials for the quantitative measurement of cortisol in serum, plasma or urine. This assay is intended for in vitro diagnostic use.
The recall was initiated because Beckman Coulter has confirmed that the listed lot numbers of Active Control ELISA reagent will result in falsely high values for patient samples due to a shift in the calibration of the assigned standard values. Beckman Coulter recommends that patient results, generated with the affected lots, be evaluated in the context of other diagnostic tests and the clinica
Beckman Coulter sent a Product Corrective Action (PCA) letter dated July 13, 2009. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use until further notice and discard remaining affected product. Customers were instructed to contact Beckman Coulter at 1-800-231-7970. Newer lots (lot number 991069 or higher) of the product have been recalibrated to restore the performance as described in the Instructions for Use (IFU). For method correlation information, please contact Technical Support at the phone number listed below. Consignees were asked to share this information with laboratory staff and retain the PCA letter for their Quality System documentation, and complete and return the enclosed Response Form within 10 days so that the firm can be assured that they received the notification. Questions regarding this notification were directed to the DSL Technical Support Center at 1-800-231-7970 in the United State and Canada, or their local Beckman Coulter Representative.
Worldwide Distribution - USA (nationwide) and Canada.
1369 Kits