FDA Recall Terminated

SIEMENS Healthineers Uroskop Omnia max Model: 10762473 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.

Recall: Z-2757-2020 · Initiated January 8, 2020

Recall

Recall Number
Z-2757-2020
Event Number
85999
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAA
Status
Terminated
Root Cause
Device Design
Initiated
January 8, 2020
Terminated
June 21, 2022
Address
40 Liberty Blvd, Malvern, PA, 19355

Description

SIEMENS Healthineers Uroskop Omnia max Model: 10762473 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.

Reason

Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis

Action

Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via XP028/19/S January 2020. Additionally, a solution to eliminate the root cause of this problem is being distributed via XP008/20/S March 2020. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. New corrective software VE10Y became available June 22, 2020, and it is being distributed via an Update Instruction XP008/20/S. This update fixes the initial reported issue and the sporadic image loss issue noted by customers who had already installed VE10W (XP029/19/S). The original Customer Safety Advisory Notice (XP028/19/S) and the information letter distributed in March 2020 remain applicable. This resolution will be provided free of charge. Your local Siemens Healthineers Service organization will contact you to arrange a date to perform this corrective action . Questions contact the service organization for an earlier appointment at 1-800-888-7436

Distribution

US Nationwide distribution.

Quantity

194 units