FDA Recall Terminated

Axiom REG implants Sterile The Axiom REG implants are used as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. There are different diameters and different lengths for each diameter.

Recall: Z-2753-2016 · Initiated June 20, 2016

Recall

Recall Number
Z-2753-2016
Event Number
74722
Firm
Anthogyr 2237 Avenue Andre Lasquin Sallanches France
FEI Number
4224
Product Code
DZE
Status
Terminated
Root Cause
Device Design
Initiated
June 20, 2016
Posted
September 8, 2016
Terminated
March 28, 2017

Description

Axiom REG implants Sterile The Axiom REG implants are used as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. There are different diameters and different lengths for each diameter.

Reason

Following a FDA inspection showing that the firm failed to validate the design and process of its implants.

Action

The firm, Anthogyr, contacted physicians by phone and sent an " URGENT: MEDICAL DEVICE RECALL" notification by mail, dated 06/20/2016, informing users of a recall due to inadequate validation of design and process of implants by the FDA. The notice described the product, problem and actions to be taken. The physicians were instructed to isolate the parts and return any Anthogyr implants from lots subject to recall to Business Area Manager, 247 SouthWest 28th road, Miami, FL 33129;and complete and return the RETURN RESPONSE form by fax to: +33 (0)4 50 93 78 60 or e.mail: [email protected] or mail: Anthogyr- Director of Quality & Regulatory Affairs- 2237 avenue- Andre Lasquin-France. The firm stated that they will not market their products in the United States until they obtain validation through 21 CFR Part 820. If you have any questions, call 404-718-9689 or email: [email protected] or [email protected].

Distribution

US Distribution to sates of: NC, NY, and OR.

Quantity

111 units