Randox Lipase Assay: Lipase Colorimetric Reagent - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.
Recall
- Recall Number
- Z-2751-2020
- Event Number
- 86039
- Firm
- RANDOX LABORATORIES, LTD. 34 Diamond Rd Crumlin Colorado Antrim United Kingdom
- FEI Number
- 3006697287
- Product Code
- CHI
- Status
- Terminated
- Root Cause
- Unknown/Undetermined by firm
- Initiated
- June 26, 2020
- Posted
- August 4, 2020
- Terminated
- October 19, 2020
Description
Randox Lipase Assay: Lipase Colorimetric Reagent - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.
Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase reagent batches.
Randox sent an urgent medical device correction notification dated June 22, 2020 explaining the current failure with select lots of Lipase Colorimetric products. The firm advised customers to scrap all effected product, keep proof of destruction, review replacements needs, review results of patients generated using affected batches and return the firm's response card.
Worldwide distribution - US Nationwide distribution including in the states of Puerto Rico and WV.
2219 kits