Affinity Four Birthing Bed Model P3700 The primary packaging is pictured below, and utilizes a metal packing brace, standard corrugated cardboard and plastic banding. When possible these beds are shipped as a quantity of two (2) together as shown with the nylon strapping. These are shipped in standard freight containers. There is no secondary packaging.
Recall
- Recall Number
- Z-2750-2016
- Event Number
- 74760
- Firm
- Hill-Rom, Inc.
- FEI Number
- 1824206
- Product Code
- KNC
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 15, 2016
- Posted
- September 8, 2016
- Terminated
- March 23, 2017
- Address
- 1069 State Road 46 E, Batesville, IN, 47006-7520
Description
Affinity Four Birthing Bed Model P3700 The primary packaging is pictured below, and utilizes a metal packing brace, standard corrugated cardboard and plastic banding. When possible these beds are shipped as a quantity of two (2) together as shown with the nylon strapping. These are shipped in standard freight containers. There is no secondary packaging.
Hill-Rom Inc. initiated a voluntary recall of the Affinity Four Birthing Beds on July 15, 2016, with letters to customers mailed certified via FedEx. Customers were asked to do the following: During use, if you encounter a situation where the CPR mechanism fails to activate, the head of the bed can still be lowered, electrically, using the siderail articulation controls. A Hill-Rom representative will be contacting each facility to make arrangements for replacing the affected head drive. Please pass this notice on to all those who need to be aware within your organization and/or to any organization where the affected devices have been transferred. If you have any questions concerning this correction, please contact Hill-Rom Technical Support at 800-445-3720.
US Distribution to the states of : CA, FL, IN, KS, LA, MI, MO, NC, NY, OK, OR, PA, TN, TX, WA, WY Foreign:BULGARIA, INDIA, UNITED ARAB EMIRATES
94 beds