Elcam Medical MANI 3G Plate BLS RAD B.BRAUN; Manifold for pressure monitoring and intravenous fluid administration. Manufactured by: Elcam Medical ACAL, Israel; Distributed by: Elcam Medical Inc., USA. Sterile
Recall
- Recall Number
- Z-2749-2015
- Event Number
- 71880
- Firm
- Elcam Medical, Inc.
- FEI Number
- 3003649401
- Product Code
- LZG
- Status
- Terminated
- Root Cause
- Packaging
- Initiated
- July 19, 2015
- Posted
- September 10, 2015
- Terminated
- September 18, 2017
- Address
- 2 University Plz, Ste 620, Hackensack, NJ, 07601-6224
Description
Elcam Medical MANI 3G Plate BLS RAD B.BRAUN; Manifold for pressure monitoring and intravenous fluid administration. Manufactured by: Elcam Medical ACAL, Israel; Distributed by: Elcam Medical Inc., USA. Sterile
Punctured blister packages were detected during the packaging process at the manufacturing site. Potential for device contamination.
Elcam Medical sent "Urgent: Stopcock Recall" notices and "Recall Return Response Forms" dated 7/19/2015 to their customers. The recall notification informed the customers of the reason for the recall and the actions they must take. Customers were instructed to cease distribution and quarantine any affected products; return to Elcam Medical the enclosed Recall Return Response Form; and notify their customers as per their recall procedures.
Texas, New Mexico and Pennsylvania
248,100