FDA Recall Terminated

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US

Recall: Z-2747-2020 · Initiated June 30, 2020

Recall

Recall Number
Z-2747-2020
Event Number
85980
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
LCP
Status
Terminated
Root Cause
Labeling design
Initiated
June 30, 2020
Terminated
November 17, 2020
Address
500 Gbc Dr, Newark, DE, 19702-2466

Description

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US

Reason

If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.

Action

Siemens Healthcare Diagnostics issues a field safety notice date June 2020 to customers

Distribution

Nationwide

Quantity

1800 units