FDA Recall Terminated

The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount

Recall: Z-2746-2016 · Initiated August 12, 2016

Recall

Recall Number
Z-2746-2016
Event Number
74915
Firm
Maquet Cardiovascular Us Sales, Llc
FEI Number
3008355164
Product Code
FQO
Status
Terminated
Root Cause
Use error
Initiated
August 12, 2016
Terminated
May 4, 2017
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount

Reason

Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.

Action

Maquet sent an "Urgent Product Recall Medical Device Correction Letter and Medical Device Correction Response Form" dated August 12, 2016 via UPS to their affected customers. Customers were advised to inspect all devices that include the fixture for damage before every use. Customers were asked to fill out and sign the enclosed Medical Device Correction Response Form and return it via email to [email protected] or fax to 1-973-909-9927. Customers with questions were advised to contact their MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383. For questions regarding this recall call 973-709-7487.